Sign In | Join Free | My himfr.com
Home > POCT Test Kit >

FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay

WWHS Biotech.Inc(exclusive marketed by Dawin)
Contact Now
    Buy cheap FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay from wholesalers
     
    Buy cheap FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay from wholesalers
    • Buy cheap FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay from wholesalers
    • Buy cheap FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay from wholesalers

    FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay

    Ask Lasest Price
    Brand Name : WWHS
    Model Number : FOB and TRF
    Certification : ISO13485
    Price : Negotiable
    Payment Terms : L/C, T/T
    Supply Ability : 50000 Kits per Week
    Delivery Time : 7 Days
    • Product Details
    • Company Profile

    FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay

    【Product name】 FOB.pdf

    Fecal occult blood (FOB) Rapid Test(Fluorescence immunoassay)


    【Intended use】

    This kit is used for quantitative determination of CEA in human whole blood, plasma and serum.

    Carcinoembryonic antigen (CEA) is a tumor associated antigen, first described in 1965 by Gold and Freedman1. It is a cell-surface glycoprotein with a molecular weight of 180-200kD, that occurs in high levels in colon epithelial cells during embryonic development. Levels of CEA are significantly lower in colon tissue of adults, but can become elevated when inflammation or tumours arise in any endodermal tissue, including the gastrointestinal tract, respiratory tract, pancreas and breast.

    An over expression of CEA protein has been detected in a variety of adenocarcinomas, including gastric, pancreatic, small intestine, colon, rectal, ovarian, breast, cervical and non-small-cell lung cancers. CEA is also expressed by epithelial cells in several non-malignant disorders, including diverticulitis, pancreatitis, inflammatory bowel disease, cirrhosis, hepatitis, bronchitis and renal failure and also in heavy smokers.

    Therefore CEA should not be regarded as a tumour-specific marker for the screening of a general population for undetected cancers. However, the determination of CEA levels provides important information about patient prognosis, recurrence of tumours after surgical removal and effectiveness of therapy.

    【Inspection principle】

    The CEA Rapid Test is a one-step chromatographic sandwich immunoassay designed for the quantitative measurement of CEA. The CEA antigen in the sample was first bound with the conjugated compound of fluorescent labeled CEA monoclonal antibody, then moved and combined with another CEA monoclonal antibody fixed on the nitrocellulose membrane, and the double antibody sandwich complex was formed at the detection line of the cellulose nitrate membrane. The quantitative detection results were obtained by NIR-1000 dry fluoroimmunoassay analyser.


    【Storage conditions and validity】

    The kit should be stored at 4℃~30℃, out of direct sunlight. It is valid for 18 months. The test card should be used within 15 minutes after unsealing under the environment of 15℃~30℃ and 20% ~ 90% relative humidity.

    The production date, batch number and expiration date are shown in the outer package of the product.

    【Applicable instruments】

    Nir-1000 dry fluorescent immunoassay analyzer produced by WWHS Biotech. Inc.


    【Procedure】

    1. Before the test, please read the instructions completely. If the test card and sample are stored in cold storage, they should be balanced at room temperature (15-30)℃ for not less than 30min before use.
    2. Start NIR-1000 dry fluoroimmunoassay analyser according to the instruction manual of the instrument, and carry out quality control verification according to the instruction manual of the instrument(Note: the reagent has been calibrated in advance, and the calibration curve parameters of each batch of reagent have been stored in the information card. The information card is inserted before use, so it is not necessary to calibrate again, and the test can be carried out only after the quality control is passed. Otherwise, the cause should be found out before testing.)
    3. Remove the test card from the aluminum foil bag and use it within 15 minutes.
    4. Place the test card on a clean horizontal table and mark it horizontally.
    5. Mix 100 µL of patient sample with 300µL of sample diluent. Apply 100 µL of diluted samples to the well of the test card.
    6. Insert the test card into NIR-1000 dry fluoroimmunoassay analyser, read and record the results at 15 minutes after addition of samples, then dispose of used test appropriately.

    【Reference interval】

    Normal persons who do not smoke: 95% had values less than 5 ng/ml. The normal reference value is 5ng/ml in this assay. It is recommended that each laboratory should establish its own normal range based on a representative sampling of the local population.


    【Limitations of methods】

    1. This kit is only used to detect human plasma/whole blood samples
    2. Due to the limitations of immunoassay methods of antigen and antibody reaction, the results cannot be used as the only basis for clinical diagnosis, but should be evaluated with all the existing clinical and experimental data.
    3. The content of triglyceride in the sample shall not exceed 15mg/ml, the content of hemoglobin shall not exceed 5mg/ml, and the content of bilirubin shall not exceed 0.5mg/ml, and the relative deviation of the test results shall not exceed ±15%.
    4. When the concentration of CEA in the sample is less than 20000ng/ml, there is no hook effect.
    5. HAMA effect was not produced when the concentration of human anti rat in the sample was less than 50ng/ml.
    6. When RF concentration in the sample is less than 2000IU/ml, the relative deviation of the test results is within ±15%.

    【Performance】

    1. Limits of detection

    No more than 1ng /ml.

    2. Accuracy

    The relative deviation from the target value is within ±15%.

    3. Precision

    The within and between assay coefficient of variations are within 15%.

    4. Linear range

    Within the linear range (1 ~ 200ng/ml), the linear correlation coefficient R≥0.990.



    Other WWHS Assay Items


    Gastrointestinal     
    cat#.Product itemSpecimenReaction TimeMeasure RangeClinical RangeItended Use
    36FOBfecal specimens10min.qualitativequalitativegastrointestinal hemorrhage.
    37TRFfecal specimens10min.qualitativequalitativegastrointestinal hemorrhage.
    38FOB+TRFfecal specimens10min.qualitativequalitativegastrointestinal hemorrhage.

    Quality FOB And TRF Rapid Test Kit With Nir-1000 25T CE Approval WWHS FIA POCT Assay for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: WWHS Biotech.Inc(exclusive marketed by Dawin)
    *Subject:
    *Message:
    Characters Remaining: (0/3000)